SPEC |
Resistance range |
10 mΩ (12.00000 mΩ display Max.,
10 nΩ resolution) to 1000 MΩ range (1200.0 MΩ display Max., 100 kΩ
resolution), 12 steps |
[LP ON]
1000 mΩ (1200.00 mΩ display Max., 10 µΩ resolution) to 1000 Ω range (1200.00
Ω display Max., 10 mΩ resolution), 4 steps |
Measurement
accuracy: ±0.006 % rdg. ±0.001 % f.s. |
Testing
current |
1 A DC to 1 µA or less [LP ON] 1
mA to 5 μA DC |
Open-terminal voltage |
20 V DC (10 kΩ range or more),
5.5 V DC max. (1000 Ω range or less) |
[LP ON]
20 mV DC max. |
Temperature
measurement |
-10.0 to 99.9 °C, accuracy:
±0.50 °C (Temperature Sensor Z2001 and RM3545 combined accuracy), -99.9 to
999.9°C (analog input) |
Measurement
speed |
FAST (2.2ms) / MED (50Hz: 21ms,
60Hz: 18ms) / SLOW1 (102ms) / SLOW2 (202ms) |
Functions |
Temperature correction,
temperature conversion, offset voltage compensation (OVC), comparator (ABS/
REF%), BIN, key-lock (OFF, menu lock, all lock), display digit count
selection function (7- digit/ 6-digit/ 5-digit), automatic power supply
frequency settings (AUTO/ 50Hz/ 60Hz), scaling, judgment sound setting, auto
hold, averaging, statistical calculations, panel store/panel load, D/A
output. |
Multiplexer |
[Only RM3545-02] Support Unit:
Z3003 (Install up to 2 units) |
Communication
interfaces |
Select from GP-IB (RM3545-01
only), RS-232C, Printer (RS-232C), or USB . Remote function, communications
monitor function, data output function, memory (50) |
Power
supply |
100 to 240 V AC, 50/60 Hz, Rated
power consumption: 40 VA |
Dimensions and mass |
215 mm (8.46 in) W × 80 mm (3.15
in) H × 306.5 mm (12.07 in) D |
[RM3545/RM3545-01]
2.5 kg (88.2 oz), [RM3545-02] 3.2 kg (35.3 oz) |
Accessories |
Power cord ×1, Clip type lead
L2101 ×1, temperature sensor Z2001 ×1, Male EXT. I/O connector ×1,
Instruction manual ×1, Application disc ×1, USB cable (A-to-B type) ×1, Spare
fuse ×1 |
| |
---|