General Specifications |
Warm up Time | | 2x the
time since last warmed up, to a maximum of 30 minutes |
System Installation | | Rear
output configuration and rack-mount kit available |
Interfaces | | IEEE-488,
RS-232, 5725A, 5205A or 5215A, 5220A, phase lock in (BNC), phase reference
out (BNC) |
Temperature Performance | Operating: | 0°C to
50°C |
Calibration: | 15°C to 35°C |
Storage: | -40°C to 75°C |
Relative Humidity | Operating: | < 80%
to 30°C, < 70% to 40°C, < 40% to 50°C |
Storage: | < 95%, non-condensing.
A power-on stabilization period of four days may be required after extended
storage at high temperature and humidity. |
Safety | | Designed
to comply with UL311; IEC 348-1978; IEC 66E (CO)4; CSA 556B |
Guard isolation | | 20 volts |
EMI/RFI | | Designed
to comply with FCC Rules Part 15, Subpart J, Class B; VDE 0871, Class B |
Line Power | | 47 to 63
Hz; + 10% allowed about selectable nominal line voltage: 100V, 110V, 115V,
120V, 200V, 220V, 230V, 240V |
| -------------- |
Maximum power: | 5700A/5720A: 300 VA;5725A,
750 VA |
Calibration Documentation | 17025 accredited report of calibration included |
Size | 5700A/5720A: | Height
17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet; width
43.2 cm (17 in), standard rack width; depth 63.0 cm (24.8 in) overall; 57.8
cm (22.7 in), rack depth |
5725A: | Height 13.3 cm (5.25 in);
width and depth same as 5700A/5720A |
| Both units project 5.2 cm,
(2 in) from rack front |
Weight | 5700A/5720A:
27 kg (62 lbs.)
5725A: 32 kg (70 lbs.)
|
| | |